Diagnostic accuracy of the parameters from ganglion cell complex map, evaluated with SD-OCT in primary open-angle glaucoma

https://doi.org/10.18008/1816-5095-2014-3-28-32

Abstract

Purpose: To evaluate the sensitivity and specificity of ganglion cell complex (GCC) parameters, obtained with optical coherence tomography (OCT) and to determine their accuracy and ability to differentiate healthy from primary open-angle glaucoma patients.

Patients and methods. 84 eyes of primary open-angle glaucoma patients and 40 eyes of healthy individuals were enrolled in the study. All of them underwent complete eye examination, including standard automated perimetry (HFA II) and OCT (RTVue-100). Avg. GCC (average GCC), Sup. GCC (superior GCC), Inf. GCC (inferior GCC), GLV (globаl loss volume), FLV (focal loss volume) and RNFL (retinal nerve fiber layer — ONH map) were measured. ROC curves
were created and sensitivity and specificity were calculated for each of these parameters.

Results.The highest sensitivity and specificity was found for GLV and the lowest for Sup. GCC. Area under the ROC curves (AUC) for GLV was found to be the largest and the smallest for Sup. GCC.

Conclusion. Parameters from GCC map have high sensitivity and specificity. Their diagnostic capability is similar, even slightly better than the one of RNFL. GLV has the highest diagnostic accuracy for primary open-angle glaucoma detection in this study.

About the Authors

B. Anguelov
Department of Ophthalmology, Medical University «Alexandrovska» Hospital
Russian Federation


K. Petrova
Department of Ophthalmology, Medical University «Alexandrovska» Hospital
Russian Federation


References

1. Greenfield DS, Bagga H, Knighton RW. Macular thickness changes in glaucomatous optic neuropathy detected using optical coherence tomography. Arch Ophthalmol 2003; 121 (1):41‑6.

2. Lederer DE, Schuman JS, Hertzmark E, et al. Analysis of macular volume in normal and glaucomatous eyes using optical coherence tomography. Am J Ophthalmol 2003; 135 (6):838‑43.

3. Medeiros FA, Zangwill LM, Bowd C, et al. Evaluation of retinal nerve fiber layer, optic nerve head, and macular thickness measurements for glaucoma detection using optical coherence tomography. Am J Ophthalmol 2005; 139 (1):44‑55.

4. Wollstein G, Schuman JS, Price LL, et al. Optical coherence tomography (OCT) macular and peripapillary retinal nerve fiber layer measurements and automated visual fields. Am J Ophthalmol 2004; 138 (2):218‑25.

5. Kendell KR, Quigley HA, Kerrigan LA, et al. Primary open-angle glaucoma is not associated with photoreceptor loss. Inv Ophthalmol Vis Sci 1995; 36 (1):200‑5.

6. Tan O, Chopra V, Lu AT, et al. Detection of macular ganglion cell loss in glaucoma by fourier-domain optical coherence tomography. Ophthalmol 2009; 116 (12):2305‑14.

7. Tan O, Li G, Lu A, et al. Advanced Imaging for Glaucoma Study Group. Mapping of macular substructures with optical coherence tomography for glaucoma diagnosis. Ophthalmol 2008; 115:949‑56.

8. Kim NR, Lee ES, Seong GJ, et al. Structure-function relationship and diagnostic value of macular ganglion cell complex measurement using Fourier-domain OCT in glaucoma. Inv Ophthalmol Vis Sci 2010; 51 (9):4646‑51.

9. Seong М, Sung КR, Choi EH, et al. Macular and peripapillary retinal nerve fiber layer measurements by Spectral domain optical coherence tomography in normaltension glaucoma. Inv Ophthalmol Vis Sci 2010; 51:1446‑52.

10. Nakano N, Hangai M, Nakanishi H, et al. Macular ganglion cell imaging in preperimetric glaucoma with speckle noise-reduced spectral domain optical coherence tomography. Ophthalmol 2011; 118 (12):2414‑26.

11. Rolle T, Briamonte C, et al. Ganglion cell complex and retinal nerve fiber layer measured by fourier-domain optical coherence tomography for early detection of structural damage in patients with preperimetric glaucoma. Clin Ophthalmol 2011; 5:961‑9.

12. Zeimer R, Asrani S, Zou S, et al. Quantitative detection of glaucomatous damage at the posterior pole by retinal thickness mapping: a pilot study. Ophthalmol 1998; 105 (2):224‑31.

13. Bagga H, Greenfield DS, Knighton RW. Macular symmetry testing for glaucoma detection. J Glaucoma 2005; 14 (5):358‑63.

14. Akashi A, Kanamori A, Nakamura M, et al. Comparative assessment for the ability of Cirrus, RTVue, and 3 D-OCT to diagnose glaucoma. Inv Ophthalmol Vis Sci 2013; 54 (7):4478‑84.

15. Takagi ST, Kita Y, Yagi F, Tomita G. Macular retinal ganglion cell complex damage in the apparently normal visual field of glaucomatous eyes with hemifield defects. J Glaucoma 2012; 21 (5):318‑25.


Review

For citations:


Anguelov B., Petrova K. Diagnostic accuracy of the parameters from ganglion cell complex map, evaluated with SD-OCT in primary open-angle glaucoma. Ophthalmology in Russia. 2014;11(3):28-32. https://doi.org/10.18008/1816-5095-2014-3-28-32

Views: 1613


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1816-5095 (Print)
ISSN 2500-0845 (Online)