Intraocular Lenses Optic Power Calculation in Extremely Short Eyes

https://doi.org/10.18008/1816-5095-2022-1-91-97

Abstract

Purpose: to choose the optimal formula for calculating the IOL optical power in patients with an axial eye length of less than 20 mm.

Patients and methods. A total of 78 patients (118 eyes) were included in the prospective study. Group I included 30 patients (52 eyes) with extremely short eyes (average axial eye length of 19.60 ± 0.42 (18.54–20.00) mm), group II consisted of 48 patients (66 eyes) with a normal axial length 22.75 ± 0.46 (22.00–23.77) mm with implantation of various monofocal IOL models. The average follow-up period was 13 months. IOL optical power was calculated using the SRK / T formula, retrospective comparison — according to the formulas Hoffer-Q, Holladay II, Olsen, Haigis and Barrett Universal II.

Results. In group I, the maximum average estimation error was determined for the formula Haigis (0.88 ± 0.35), then for the formula Olsen, Barrett Universal II, Kane, SRK / T, Holladay 2 and Hoffer-Q (0.51 ± 0.12, 0.16 ± 0.38, 0.13 ± 0.28, 0.10 ± 0.59, 0.05 ± 0.54 and –0.12 ± 0.42, respectively). Similar data were obtained for the average absolute error — for the formulas Haigis, Olsen, Barrett Universal II, SRK / T, Holladay 2, Hoffer-Q and Kane, it was 0.85 ± 0.31, 0.78 ± 0.25, 0.21 ± 0.10, 0.79 ± 0.23, 0.73 ± 0.24, 0.19 ± 0.08 and 0.17 ± 0.06, respectively. When comparing the formulas, significant differences were found for the formulas Hoffer-Q, Barrett Universal II and Kane in comparison with the formulas Haigis, Olsen, SRK / T and Holladay II (p < 0.05) in all cases, respectively, which indicates the advantage of these formulas for the group patients with extremely short eyes. In group II, there were no significant differences between the studied formulas (p > 0.05).

Conclusion. This paper presents an analysis of our own data on the effectiveness of six formulas for calculating the IOL optical power in extremely short (less than 20 mm) eyes in comparison with the normal axial length. The advantage of the Hoffer-Q, Barrett Universal II and Kane formulas over Haigis, Holladay 2, Olsen, and SRK / T is shown. To determine the exact indications for using these formulas, further studies are necessary taking into account the anterior chamber depth and a lesser degree of hyperopia.

About the Authors

К. B. Pershin
“Eximer” Eye Center; Academy of Postgraduate Education of the Federal Medical-Biological Agency
Russian Federation

Pershin Kirill B., MD, Professor, medical director

Marksistskaya str., 3/1, Moscow, 109147; Volokolamskoe highway, 91, Moscow, 125371



N. F. Pashinova
“Eximer” Eye Center; Academy of Postgraduate Education of the Federal Medical-Biological Agency
Russian Federation

Pashinova Nadezhda F., MD, head doctor

Marksistskaya str., 3/1, Moscow, 109147; Volokolamskoe highway, 91, Moscow, 125371



I. A. Likh
“Eximer” Eye Center
Russian Federation

Likh Ivan A., ophthalmologist

Family Shamshinikh str., 58, Novosibirsk, 630005 



А. Yu. Tsygankov
“Eximer” Eye Center
Russian Federation

Tsygankov Alexander Yu., PhD, scientific advisor

Marksistskaya str., 3/1, Moscow, 109147



S. L. Legkikh
“Eximer” Eye Center
Russian Federation

Legkikh Sergei L., PhD, ophthalmologist

Marksistskaya str., 3/1, Moscow, 109147



References

1. Hahn U., Krummenauer F., Kolbl B., Neuhann T., Schayan-Araghi K., Schmickler S., von Wolff K., Weindler J., Will T., Neuhann I. Determination of valid benchmarks for outcome indicators in cataract surgery: a multicenter, prospective cohort trial. Ophthalmology. 2011;118:2105–2112. DOI: 10.1016/j.ophtha.2011.05.011

2. Pershin K.B., Pashinova N.F., Tsygankov A.Yu., Legkih S.L., Lih I.A. Biometry in lOL power calculations as a factor of successive cataract surgery. Cataract and refractive surgery = Kataraktal’naya i refraktsionnaya khirurgiya.2016;16(2):15–22 (In Russ.).

3. Hoffer K.J. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–712. DOI: 10.1016/s0886-3350(13)80338-0

4. Olsen T., Thim K., Corydon L. Accuracy of the newer generation intraocular lens power calculation formulas in long and short eyes. J Cataract Refract Surg. 1991;17:187–193. DOI: 10.1016/s0886-3350(13)80249-0

5. Steijns D., Bijlsma W.R., Van der Lelij A. Cataract surgery in patients with nanophthalmos. Ophthalmology. 2013;120:266–270. DOI: 10.1016/j.ophtha.2012.07.082

6. Jung K.I., Yang J.W., Lee Y.C., Kim S.Y. Cataract surgery in eyes with nanophthalmos and relative anterior microphthalmos. Am J Ophthalmol. 2012;153:1161–1168 e1. DOI: 10.1016/j.ajo.2011.12.006

7. Fenzl R.E., Gills J.P., Cherchio M. Refractive and visual outcome of hyperopic cataract cases operated on before and after implementation of the Holladay II formula. Ophthalmology. 1998;105:1759–1764. DOI: 10.1016/s0161-6420(98)99050-9

8. Fedorov S.N., Kolinko A.I. A method of calculating the optical power of the intraocular lens. Annals of Ophthalmology = Vestnik oftal’mologii. 1967;80:27–31 (In Russ.).

9. Hoffer K.J. Clinical results using the Holladay 2 intraocular lens power formula. J Cataract Refract Surg. 2000;26:1233–1237. DOI: 10.1016/s0886-3350(00)00376-x

10. Trivedi R.H., Wilson M.E., Reardon W. Accuracy of the Holladay 2 intraocular lens formula for pediatric eyes in the absence of preoperative refraction. J Cataract Refract Surg. 2011;37:1239–1243. DOI: 10.1016/j.jcrs.2011.01.021

11. Carifi G., Aiello F., Zygoura V., Kopsachilis N., Maurino V. Accuracy of the refractive prediction determined by multiple currently available intraocular lens power calculation formulas in small eyes. Am J Ophthalmol. 2015;159:577–583. DOI: 10.1016/j.ajo.2014.11.036

12. Srivannaboon S., Chirapapaisan C., Chirapapaisan N., Lertsuwanroj B., Chongchareon M. Accuracy of Holladay 2 formula using IOL Master parameters in the absence of lens thickness value. Graefe’s Arch Clin Exper Ophthalmol. 2013;251:2563–2567. DOI: 10.1007/s00417-013-2439-8

13. Terzi E., Wang L., Kohnen T. Accuracy of modern intraocular lens power calculation formulas in refractive lens exchange for high myopia and high hyperopia. J Cataract Refract Surg. 2009;35:1181–1189. DOI: 10.1016/j.jcrs.2009.02.026

14. Kane J.X., Van Heerden A., Atik A., Petsoglou C. Intraocular lens power formula accuracy: comparison of 7 formulas. J Cataract Refract Surg. 2016;42:1490–1500. DOI: 10.1016/j.jcrs.2016.07.021

15. MacLaren R.E., Natkunarajah M., Riaz Y., Bourne R.R., Restori M., Allan B.D. Biometry and formula accuracy with intraocular lenses used for cataract surgery in extreme hyperopia. Am J Ophthalmol. 2007;143:920–931. DOI: 10.1016/j.ajo.2007.02.043

16. Moschos M.M., Chatziralli I.P., Koutsandrea C. Intraocular lens power calculation in eyes with short axial length. Indian J Ophthalmol. 2014;62:692–694. DOI: 10.4103/0301-4738.129791

17. Day A.C., Foster P.J., Stevens J.D. Accuracy of intraocular lens power calculations in eyes with axial length <22.00 mm. Clin Experiment Ophthalmol. 2012;40:855–862. DOI: 10.1111/j.1442-9071.2012.02810.x

18. Gavin E.A., Hammond C.J. Intraocular lens power calculation in short eyes. Eye. 2008;22:935–938. DOI: 10.1038/sj.eye.6702774

19. Cooke D.L, Cooke T.L. Comparison of 9 intraocular lens power calculation formulas. J Cataract Refract Surg. 2016;42:1157–1164. DOI: 10.1016/j.jcrs.2016.06.029

20. Narvaez J., Zimmerman G., Stulting R.D., Chang D.H. Accuracy of intraocular lens power prediction using the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas. J Cataract Refract Surg. 2006;32:2050–2953. DOI: 10.1016/j.jcrs.2006.09.009

21. Bat’kov E.N., Pashtaev N.P., Mihajlova V.I. Calculation of intraocular lens power in surgical treatment of extreme hyperopia. Annals of Ophthalmology = Vestnik oftal’mologii. 2019;135(1):21–27 (In Russ.). DOI: 10.17116/oftalma201913501121

22. Rocha-de-Lossada C., Colmenero-Reina E., Flikier D., Castro-Alonso F.J., Rodriguez-Raton A., García-Madrona J.L., Peraza-Nieves J., Sánchez-González J.M. Intraocular lens power calculation formula accuracy: Comparison of 12 formulas for a trifocal hydrophilic intraocular lens. Eur J Ophthalmol. 2020:1120672120980690. DOI: 10.1177/1120672120980690. Epub ahead of print.

23. Darcy K., Gunn D., Tavassoli S., Sparrow J., Kane J.X. Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service. J Cataract Refract Surg. 2020;46(1):2–7. DOI: 10.1016/j.jcrs.2019.08.014


Review

For citations:


Pershin К.B., Pashinova N.F., Likh I.A., Tsygankov А.Yu., Legkikh S.L. Intraocular Lenses Optic Power Calculation in Extremely Short Eyes. Ophthalmology in Russia. 2022;19(1):91-97. (In Russ.) https://doi.org/10.18008/1816-5095-2022-1-91-97

Views: 727


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1816-5095 (Print)
ISSN 2500-0845 (Online)