Purpose. To optimize a-constants for intraocular lenses (IOL) on the Clareon® (Alcon, USA) platform. 

Material and methods. The study included 206 patients (206 eyes) divided into 4 groups depending on IOL implanted model: the 1st — SY60WF (= 71); the 2nd — Clareon Toric CNW0T2-9 (= 76); the 3rd — Clareon PanOptix CNWTT0-Т6 (= 32); the 4th — Clareon Vivity CNWET0-Т6 (= 27). Optical biometry was performed by IOL-Master 700 (Carl Zeiss, Germany). For toric IOL calculation keratotopography by CASIA Corneal/Anterior Segment OCT SS-1000 (Tomey, Japan) was performed. A month after phacoemulsification spherical equivalent of refraction was evaluated. Mean IOL calculation error (ME) was used to calculation accuracy assessment. A-constant optimization performed by ME zeroing. 

Results. ME was 0.32 ± 0.42 D for Clareon SY60WF, 0.31 ± 0.36 D for Clareon Toric CNW0T2-9, 0.38 ± 0.34 D for Clareon PanOptix CNWTT0-Т6 and 0.10 ± 0.34 D for Clareon Vivity CNWET0-Т6. ME zeroing allows to define real a-constants for IOLs: 119.38 for Clareon SY60WF, 119.46 for Clareon Toric CNW0T2-9, 119.54 for Clareon PanOptix CNWTT0-Т6 and 119.30 for Clareon Vivity CNWET0-Т6. 

Conclusion. Using of proposed a-constants for IOLs on Clareon® platform could help with target refraction achievement and to avoid a hyperopic refraction shift.

About the Authors

D. F. Belov
Saint-Petersburg Multifield Hospital no. 2; Saint Petersburg State University
Russian Federation

Belov Dmitrii F., PhD, chief of Ophthalmological Department no. 1 

Uchebniy lane, 5, Saint-Petersburg, 194354;
University emb., 7/9, Saint-Petersburg, 199034



T. Yu. Kulikaeva
Saint-Petersburg Multifield Hospital no. 2
Russian Federation

Kulikaeva Tatiana Yu., ophthalmologist of Ophthalmological Department no. 1 

Uchebniy lane, 5, Saint-Petersburg, 194354



V. P. Nikolaenko
Saint-Petersburg Multifield Hospital no. 2; Saint Petersburg State University
Russian Federation

Nikolaenko Vadim P., MD, рrofessor the of Otolaryngology and Ophthalmology Department, deputy chief physician of Ophthalmology

Uchebniy lane, 5, Saint-Petersburg, 194354;
University emb., 7/9, Saint-Petersburg, 199034

 



References

1. Belov DF, Nikolaenko VP, Dmitrieva DE. Accuracy of 10 Intraocular Lens Power Calculation Formulas. Ophthalmology In Russia. 2025;22(1):29–34 (In Russ.). doi: 10.18008/1816‑5095‑2025‑1‑29‑34.

2. Melles RB, Kane JX, Olsen T, Chang WJ. Update on Intraocular Lens Calculation Formulas. Ophthalmology. 2019 Sep;126(9):1334–1335. doi: 10.1016/j.ophtha.2019.04.011.

3. Rbfcalculator.com. URL: https://rbfcalculator.com/online/index.html (accessed: 20.11.2025).

4. Iolformula.com. https://www.iolformula.com/ (accessed: 20.11.2025).

5. Evoiolcalculator.com. https://www.evoiolcalculator.com/ (accessed: 20.11.2025).

6. Or L, Jacques A, Barrett GD. Autorefraction as an Objective Method to Evaluate Accuracy of Intraocular Lens Calculation Formulas. J Refract Surg. 2022 Sep;38(9):580– 586. doi: 10.3928/1081597X‑20220715‑01.

7. Belov DF, Nikolaenko VP, Kovaleva VV. Evaluation and Refractive Results Comparison of MIOLSOFT213 IOL Implantation with Foreign Models. Ophthalmology In Russia. 2024;21(2):289–295 (In Russ.). doi: 10.18008/1816-5095-2024-2-289-295

8. Bellucci C, Mora P, Tedesco SA. Automated and subjective refraction with monofocal, multifocal, and EDOF intraocular lenses: review. J Cataract Refract Surg. 2023 Jun 1;49(6):642–648. doi: 10.1097/j.jcrs.0000000000001186.

9. Hoffer KJ, Savini G. Update on Intraocular Lens Power Calculation Study Protocols: The Better Way to Design and Report Clinical Trials. Ophthalmology. 2021 Nov;128(11):e115–e120. doi: 10.1016/j.ophtha.2020.07.005.

10. Abulafia A, Barrett GD, Koch DD. Protocols for Studies of Intraocular Lens Formula Accuracy. Am J Ophthalmol. 2016 Apr;164:149–50. doi: 10.1016/j.ajo.2016.01.010.

11. Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990 May;16(3):333–340. doi: 10.1016/s0886‑3350(13)80705‑5.

12. Zhang C, Dai G, Pazo EE.Accuracy of intraocular lens calculation formulas in cataract patients with steep corneal curvature. PLoS One. 2020 Nov 20;15(11):e0241630. doi: 10.1371/journal.pone.0241630.

13. Ning X, Yang Y, Yan H, Zhang J. Anterior chamber depth — a predictor of refractive outcomes after age‑related cataract surgery. BMC Ophthalmol. 2019 Jun 25;19(1):134. doi: 10.1186/s12886‑019‑1144‑8.

14. Hoffer KJ, Savini G. IOL power calculation in short and long eyes. Asia Pac J Ophthalmol (Phila). Jul‑Aug 2017;6(4):330–331. doi: 10.22608/APO.2017338.

15. Meng J, Fang Y, Lian J. Visual and patient‑reported outcomes of a diffractive trifocal intraocular lens in highly myopic eyes: a prospective multicenter study. Eye Vis (Lond). 2023 Apr 6;10(1):19. doi: 10.1186/s40662‑023‑00336‑3.

16. Shen J, Cai L, Zhuo B. Binocular Visual Outcomes Comparison of Two Trifocal Intraocular Lenses in High‑Myopic Cataract Patients: A 1‑Year Multicenter Study. Am J Ophthalmol. 2023 Oct;254:1–10. doi: 10.1016/j.ajo.2023.03.015.

17. Belov DF, Nikolaenko VP. Alternative method of intraocular lens power calculation in eyes with short axial length. Russian Annals of Ophthalmology. 2022;138(3):24– 28 (In Russ.). doi: 10.17116/oftalma202213803124.

18. Batkov EN, Pashtayev NP, Mikhaylova VI. Calculation of intraocular lens power in surgical treatment of extreme hyperopia. The Russian Annals of Ophthalmology 2019;135(1):21–27 (In Russ.). doi: 10.17116/oftalma201913501121.

19. Tuuminen R, Jeon S, Jung BJ, Moon K. Prognostic factors for dysphotopsia and spectacle‑independent visual function after implantation of non‑diffractive extending focus intraocular lenses. Acta Ophthalmol. 2025 Jul 15. doi: 10.1111/aos.17560.

20. Chaperon M, Chehab H, Levron A.Preliminary refractive results of the Clareon® Intraocular Lens (IOL): about 150 cases. Invest. Ophthalmol. Vis. Sci. 2019;60(9):3712.

21. Choi S, Choi YS, Nam DJ, Na S. One‑year outcomes of a trifocal intraocular lens with glistening‑free hydrophobic acrylic using the mix‑and‑match technique. BMC Ophthalmol. 2025 Jul 1;25(1):370. doi: 10.1186/s12886‑025‑04198‑2.

22. Belov DF, Potemkin VV, Nikolaenko VP. Optimization of intraocular lens power calculation in pseudoexfoliation syndrome. The Russian Annals of Ophthalmology. 2021;137(4):38–42 (In Russ.). doi: 10.17116/oftalma202113704138.

23. Iolcon.org. URL: https://iolcon.org/lensesTable.php?manufacturer%5B%5D=Alcon&action=search (accessed: 20.11.2025)


Review

For citations:


Belov D.F., Kulikaeva T.Yu., Nikolaenko V.P. A-constants Optimization for Intraocular Lenses on the Clareon® platform. Ophthalmology in Russia. 2026;23(1):53-58. (In Russ.) https://doi.org/10.18008/1816-5095-2026-1-53-58

Views: 383

JATS XML


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1816-5095 (Print)
ISSN 2500-0845 (Online)